134
|
4 Measurement of Biosignals and Analog Signal Processing
Fig. 4.41: RLC-Potency-Filter-Bandstop 2. orders (left) and associated frequency response (calcula-
tion with LTspice, right).
Fig. 4.42: ECG with 50 Hz mains hum before and after filtering with a potency-band block 2nd order.
4.5.1.3 Chebyshev Filter
In a Chebyshev-filter, for the characteristic function K(Ω) a Chebyshev-polynomial
indexChebyshev-polynomial Tn(Ω) weighted by a factor ϵ is used.
K(Ω) = ϵ ⋅Tn(Ω)
(4.44)
is chosen, with
Tn(Ω) = cos(n ⋅arccos(Ω))
for |Ω| < 1 Tn(Ω)= cosh(n ⋅arccosh(Ω))
for |Ω| > 1 .
(4.45)
The value n in turn determines the order of filtering. Examples for Tn(Ω) are:
T1(Ω) = Ω(like power filter) ,
T2(Ω) = −1 + 2 Ω2 ,
T3(Ω) = −3 Ω+ 4 Ω3 ,
T4(Ω) = 1 −8 Ω2 + 8 Ω4 ,
T5(Ω) = 5 Ω−20 Ω3 + 16 Ω5 .
(4.46)